A0 value
The A0 value describes the effectiveness of thermal disinfection as a combination of temperature and exposure time. It allows different programmes to be compared: a higher temperature reaches the same A0 value in a shorter time.
The German Society for Hospital Hygiene (DGKH) names two reference values:
- A0 600 – corresponds to 90 °C for 1 minute. Effective against bacteria, mycobacteria, fungi and heat-sensitive viruses.
- A0 3000 – corresponds to 90 °C for 5 minutes. Required to also inactivate heat-resistant viruses such as hepatitis B. DGKH generally recommends this value for programmes reprocessing surgical instruments.
Which value applies to your instruments follows from the risk assessment; for surgical instruments it is A0 3000 according to DGKH. Whether it is reached is demonstrated by validation with temperature measurements in the device. In the disinfection phase the HYG series heats up to 93 °C and reaches an A0 value of over 3000. Only then is disinfection rated as successful. More: HYG 3 · HYG 5
Related terms: Thermal disinfector · Validation process · Semi-critical B