Bowie-Dick-Test
The Bowie-Dick test is a steam penetration test for sterilisers with fractionated pre-vacuum, i.e. for the class B autoclave. It shows whether air is completely removed from the chamber and steam reaches the load everywhere.
A test pack or test device with an indicator runs through a special test programme. A uniform colour change of the indicator shows adequate steam penetration.
In practices the test is usually performed every working day before the first batch, often together with the vacuum test. For hollow instruments, a helix test is often used. The steriliser manufacturer specifies how often to test. The result belongs in the batch documentation.
Related terms: Vacuum test · Sterilisation cycle · Class B autoclave