Your Cart is Empty!
Back to overview

DIN EN ISO 11607-2

ISO 11607 (DIN EN ISO 11607) is the standard for packaging of terminally sterilised medical devices. It applies in Germany and throughout the EU and has two parts:

  • Part 1: requirements for materials, sterile barrier systems and packaging systems
  • Part 2: validation requirements for forming, sealing and assembly processes

For practices, Part 2 is particularly relevant: anyone who seals instruments into clear-view pouches with a sealer must validate this sealing process. This includes setting and regularly checking sealing temperature, contact pressure and speed, as well as testing the seal, e.g. by a seal strength test.

Related terms: Packaging system · Seal strength testing · Sterile barrier system