Drying

Drying is the last step in a WD programme. Residual moisture on and in instruments must be removed because germs can multiply in moisture and wet instruments should not be packed and sterilised.

Hollow instruments are particularly critical: water easily remains in lumens and joints. That is why validation also checks whether the instruments are dry at the end of the programme.

If residual moisture remains after the programme, it is removed with medical compressed air or lint-free cloths.

The HYG series dries the instruments at the end of the cycle using the residual heat of the load. The door should be opened within 15 minutes of the end of the programme; if it stays closed longer, water stains may form or instruments may remain damp. For transmission instruments, many manufacturers additionally recommend drying the spray, air and water channels with medical compressed air. More: HYG series thermal disinfectors

Related terms: Hollow instruments · Sterilisation · Validation process