Performance qualification (LQ)
Performance qualification is the provision and documentation of evidence that the equipment, as installed and operated in accordance with the operating procedures, continuously performs in accordance with the predetermined criteria, thereby obtaining products that meet its specifications. To demonstrate short-term reproducibility, a loading configuration (loading trolley + program) is tested three times.
During the performance qualification, the defined cleaning and disinfection programs for typical company loads (reference loads) are checked and the results are documented. Test specimens are used to check the basic cleaning performance. Test specimens are Crile clamps, which are clearly contaminated with test contamination, and instruments that are actually contaminated, so that a reference to practical loading can be established.
Acceptance criteria for the cleaning performance according to the validation guidelines (the test instrument must be visually clean):
Value – Requirement for residual protein quantities in µS/cm
Guideline – Less 80 µg protein
Warning level – 80 µg to 150 µg
Limit – Greater than 150 µg