Thermal disinfection
Thermal disinfection kills or inactivates pathogens with moist heat, i.e. hot water at a defined temperature for a defined time. It is the core step in the thermal disinfector and works without chemical disinfectants.
How effective a programme is, is described by the A0 value. An A0 value of 600 corresponds to 90 °C for 1 minute; A0 3000 corresponds to 90 °C for 5 minutes and also inactivates heat-resistant viruses such as hepatitis B.
Compared with chemical or chemo-thermal disinfection, the process has two advantages: it leaves no disinfectant residues, and temperature and time can be measured and documented exactly. The prerequisite is that the instruments are thermostable and have been thoroughly cleaned beforehand.
The HYG series thermal disinfectors disinfect thermally in a fixed programme. More: HYG series thermal disinfectors
Related terms: A0 value · Exposure time · WD-E