Urology

Urology deals with diseases of the urinary organs in women and men and of the male reproductive organs.

Besides standard instruments such as clamps, tweezers and scissors, urological practices reprocess rigid hollow instruments, for example thermostable endoscopes such as cystoscopes or rectoscopes. The manufacturer’s instructions are decisive: if an instrument is approved for thermal reprocessing, it belongs in a normal thermal disinfector. Only heat-sensitive, mostly flexible endoscopes must be reprocessed chemo-thermally in a WD-E.

Many urological instruments are long and have narrow lumens. They must be flushed through suitable adapters so that the inner surfaces are also cleaned and disinfected.

The “Urology” package for the HYG 5 includes the Flexio square injection rail HYG 5 (12 adapters, one or two rails) and a medical silicone hose for connecting hollow instruments. More: HYG 5 – Urology