Validation Guideline
The validation guideline is the “Guideline compiled by DGKH, DGSV and AKI for the validation and routine monitoring of automated cleaning and thermal disinfection processes for medical devices”. It is published by the German Society for Hospital Hygiene, the DGSV and the Instrument Reprocessing Working Group (AKI).
It describes how operators demonstrate that their washer-disinfector reprocesses reproducibly clean and safe. It is based on the ISO 15883 standard series. Validation is divided into:
- Installation qualification (IQ): is the device correctly set up and connected?
- Operational qualification (OQ): does it work as specified by the manufacturer?
- Performance qualification (PQ): does it achieve the required result with the practice’s loads?
In addition, there are routine checks and repeated performance qualification at set intervals. More: HYG series thermal disinfectors