Validation process

Validation is the documented proof that a reprocessing procedure reliably delivers the intended result under the conditions of your practice – clean, disinfected and dry instruments. Under section 8 of the German MPBetreibV, medical devices used low-germ or sterile must be reprocessed with suitable validated procedures.

For mechanical cleaning and disinfection processes, the guideline of DGKH, DGSV and AKI describes the procedure. Validation consists of three steps:

  1. Installation qualification: Is the device correctly set up and connected – are water, power and drainage right?
  2. Operational qualification: Does the device work technically as specified by the manufacturer?
  3. Performance qualification: Are typical practice loads demonstrably cleaned and disinfected? This is tested with test soils, temperature measurements and protein tests.

Validation is carried out by a competent, independent body. IC Medical deliberately does not validate itself: we advise you on validation, help you choose an accredited partner laboratory and arrange the appointment when you buy the device – and later the revalidations too, which we remind you of in good time. This keeps the testing independent. More: Appointment