Practice inspection: what is checked in instrument reprocessing

Geschrieben von icmedical

Zuletzt aktualisiert: 18. October 2026

For many practices, the announcement of an inspection makes the pulse rise. Yet an inspection can be planned well: the authorities check reprocessing according to a comprehensible pattern. If you know what matters and have the documents at hand, you can face the inspection calmly. This guide refers to the situation in Germany.

Who inspects – and why?

The reprocessing of medical devices is monitored from two sides: by the public health office on the basis of the Infection Protection Act and by medical device surveillance on the basis of the Medical Devices Operator Ordinance. In Baden-Württemberg the regional councils are responsible, in other federal states other state authorities. Dental practices are also familiar with hygiene inspections as part of quality assurance.

In all cases, the benchmark is the joint recommendation of KRINKO and BfArM on the reprocessing of medical devices.

What is checked in reprocessing

1. Organisation and documents

  • current, signed hygiene plan that matches the actual workflows
  • risk assessment and classification of the medical devices (non-critical, semi-critical A/B, critical A/B) – and matching reprocessing procedures
  • standard operating procedures for every reprocessing step
  • proof of competence of the staff who reprocess
  • instruction records for the devices used

2. Devices and procedures

  • validation reports for the thermal disinfector and steriliser – including valid revalidation
  • maintenance records according to the manufacturer’s specifications
  • inventory and, where required, medical device log
  • seal strength testing of packaging, routine checks of the devices

3. Batch documentation and release

  • documentation of every batch with the process parameters
  • documented release by an authorised person
  • labelling of sterile-packed instruments (sterilisation date, storage period)

4. Rooms and workflows on site

  • clear separation of unclean and clean areas in the reprocessing room
  • proper storage of instruments and process chemicals
  • Inspectors often ask to be shown the workflows – it is not enough if they are only correct on paper.

The most common deficiencies

  1. risk classification missing or not matching the procedure
  2. hygiene plan outdated or not adapted to the practice
  3. proof of competence missing
  4. validation missing or revalidation overdue
  5. gaps in batch documentation or release
  6. no clear separation of clean and unclean

Checklist: how to prepare

  • ☐ Hygiene plan checked, updated and signed
  • ☐ Risk classification of all medical devices documented
  • ☐ Standard operating procedures complete and up to date
  • ☐ Validation reports for thermal disinfector and steriliser available, revalidation not overdue
  • ☐ Maintenance records available
  • ☐ Batch documentation and releases complete
  • ☐ Proof of competence and instruction records of everyone involved at hand
  • ☐ Reprocessing room tidy, clean and unclean visibly separated
  • ☐ Team knows who answers which questions and demonstrates workflows

How a thermal disinfector makes the inspection easier

Machine reprocessing in a thermal disinfector is standardised, reproducible and easy to validate. The thermal disinfectors HYG 3 and HYG 5 support you:

  • One programme: no programme selection, no operating errors – which simplifies work instructions and user instruction.
  • Automatic batch recording with the IC Medical Printer Software for complete documentation.
  • Tabs instead of liquid chemicals: no canisters and dosing pumps that need to be stored, changed and checked.
  • Initial validation in the complete offer and reminders for maintenance and revalidation.
  • Work instructions: we support you with work instructions for the HYG devices.

Frequently asked questions about practice inspections

Who carries out practice inspections of reprocessing?

This is regulated differently in each German federal state. Responsible are the public health office (Infection Protection Act) and the state authority responsible for medical device surveillance – in Baden-Württemberg the regional councils (Regierungspräsidien), in other states for example district governments or state offices.

Is an inspection announced in advance?

Routine inspections are usually announced a few weeks in advance, often together with a questionnaire and a list of documents. Unannounced inspections are legally possible, for example when there is a specific reason.

Which documents should be ready for an inspection?

Among others: the current hygiene plan, the risk assessment and classification of the medical devices, standard operating procedures, the validation reports for the thermal disinfector and steriliser, maintenance records, batch documentation with releases, instruction records and the proof of competence of the staff.

What are the most common deficiencies in reprocessing?

Frequently objected to are missing or incorrect risk classifications, outdated hygiene plans, missing proof of competence, missing or expired validations and insufficient separation of unclean and clean areas.

What happens if deficiencies are found?

As a rule, you receive a report with a deadline for remedying them. In the case of serious or repeated deficiencies, requirements, fines and, in extreme cases, a ban on reprocessing are possible. Follow-up inspections may incur additional fees.

Conclusion

An inspection checks whether your reprocessing is demonstrably safe. With up-to-date documents, validated procedures and complete batch documentation, you are well prepared. If you want to modernise your reprocessing or have questions about validation: contact us – by phone on +49 7181 70600.

Note: This guide provides a general overview of the situation in Germany (as of October 2026) and does not constitute legal advice. Procedures and responsibilities differ between federal states; the requirements of your competent authority are decisive.