Thermal disinfector HYG 3 – specialization in gynecology

Read more about hygiene in gynaecology in our hygiene lexicon: Gynaecology

Zusatzleistungen
Installation package

We come to you, install your thermodisinfector and instruct you on how to use it.

WaterDemin 550

Fully demineralized water sustainably ensures the maximum service life of your devices and instruments.

Initial validation

We arrange the initial validation through one of our accredited partner laboratories and advise you on your processes in advance. We also arrange revalidations and remind you in good time when they are due.

Send Inquiry
			
Grundkorb (Unterkorb) des HYG 3 / HYG 5
Basic basket (item number 120969)

Dimensions (HxWxD) 6.7 x 35 x 40.7 cm

With this basic basket you can easily insert gynecological instruments. The basis is our Flexio organization system, which allows a load of up to 5 kg.


Flexio Universalhalter für 8 Spekula oder 21 Zangen
2x specula holder (item number 105034)

The speculum holder can hold up to 8 speculums ‘Cusco style’ or 5 speculums ‘Kristeller style’.

At a glance again: The HYG 3 thermodisinfector with the GYN package offers the potential to

  • 3 x speculum holders in the basic basket: up to 24 speculums ‘Cusco style’ or
  • 3 x speculum holders in the basic basket: up to 15 speculums ‘style Kristeller’

processed safely and gently in just one cycle.

As an option, we also offer the instrument holder (item number 105011) for the preparation of other instruments.


Flexio Instrumentenhalter-Einsatz
1x instrument holder (item number 105011)

(HxWxL) 9.4 x 11 x 36.5 cm

Number of grids = 96 pieces (10×10 mm) + 8 pieces 25×25 mm

Tweezers, probes, eyelet holders, curettes and other instruments can be easily picked up while standing. There are also 8 additional, larger compartments available. You can also use these for scissors, clamps, pliers, needle holders, etc. Up to 96 standing instruments can be installed in the HYG 3.


Chargendokumentation des HYG: Prozessprotokoll am PC
1x documentation software IC Medical Printer (item number 106808)

The IC Medical Printer documentation software included in the scope of delivery ensures legally compliant process documentation with all prescribed parameters. For greater convenience, batch release takes place directly on the device display of the thermal disinfector. The IC Medical Printer software is installed on your practice computer or network. The data can be easily read and archived using the included USB stick or via a direct network connection. Complete. We will integrate it into your practice software free of charge. Interfaces to Segosoft, DIOS and DOCma are available.


Symbol Flexio Ordnungssystem
Info: Flexio filing system

Thanks to our Flexio organization system, all equipment packages can be individually adapted to your practice needs. With over 50 other items such as baskets, adapters, holders and inserts, we will find the right solution for your requirements. Simply contact us: vertrieb@icmedical.de / Telephone 07181 70600.


Pumpe des HYG Thermodesinfektors
Innovative for more safety

We have developed innovative pump technology and a new water supply system for our HYG 3. Both ensure thorough cleaning and disinfection of your instruments.


Symbol Programmlaufzeit
Short program time for more efficiency

Our thermal disinfector HYG 3 processes even demanding batches reliably, quickly and safely. The construction is stable and powerful. A single loading level keeps loading simple and operation smooth. Even when there are high or changing instrument requirements, the HYG 3 delivers full power. For reprocessing hollow instruments, we recommend the HYG 5.


Thermo Tabs für die Tab-Technologie der HYG-Serie – Dosierung ohne Flüssigchemie
Process chemicals - easy to store, perfectly dosed and space-saving

Our proven and popular Oxides ThermoTabs standardize and make your processing process much easier. The dosage and concentration of the cleaning components are consistent and perfectly coordinated. The storage effort is reduced enormously and incorrect dosages are practically eliminated. Simply.


Instrumente nach der Aufbereitung im HYG Thermodesinfektor
IC HotDrying process

With the IC HotDrying process, instruments dry using the residual heat of the programme. Simply open the door within 15 minutes after the programme ends. This prevents water stains and damp instruments.


Chargenfreigabe direkt am Display des HYG Thermodesinfektors
Batch release directly on the display for more convenience

The visual inspection is mandatory. With the HYG 3, we make batch release even easier for you: you can then release it directly on the device display. Our software transfers the release to the batch record at the same time.


Chargendokumentation des HYG: Prozessprotokoll am PC
Interfaces for legally compliant documentation

Safe instrument reprocessing includes legally compliant process documentation with all prescribed parameters. The HYG 3 offers you interfaces to easily read and archive the data via a network connection or USB stick. We provide our IC Printer documentation software for this purpose. In addition, the documentation can be integrated into your existing practice software via other interfaces (Segosoft, DOGma (DIOS)).


WaterDemin 550 zur Vollentsalzung für Thermodesinfektoren
WaterDemin 550 for valid water quality

Fully demineralized water sustainably ensures the maximum service life of your devices and instruments. For full desalination we offer the WaterDemin 550.
Read more about the topic of device-friendly water treatment: Everything you need to know about water treatment.
The IC Touch Control Display of the HYG 3 shows you immediately when the performance of your VE cartridge is exhausted. Alternatively, your batch record provides information about this.


Wasser- und Stromanschluss des HYG Thermodesinfektors
Installation requirements

To operate the thermal disinfector HYG 3, you only need:

  • Power supply 220–240 V; 50/60 Hz; 16 A, separately fused
  • Cold water connection 3/4 inch
  • Drain connection DN19
  • Demineralised water

Type designation HYG 3
Article number 120 003
Programme duration approx. 70 minutes (one programme for all loads)
Dimensions countertop unit (H x W x D) 62.5 x 43 x 50 cm (with handle + 4.5 cm)
Dimensions undercounter unit incl. base (H x W x D) 82 x 43 x 50 cm (with handle + 4.5 cm)
Door projection (A) 27 cm
Wash chamber, max. load space (W x H x D) 21.5 x 34.6 x 40.3 cm
Weight (empty) 28 kg
Floor load approx. 114 N/m²
Wash chamber volume 96 litres
Electrical connection 220–240 V; 50/60 Hz; 16 A, separately fused; 30 mA RCD
Max. sound power (cleaning) approx. 59 dB
Heat emission < 40 °C
Ambient temperature 15–35 °C
Relative humidity 30–60 %
Max. installation altitude 2000 m
Installation category 2
Air pressure 75–106 kPa
Cold water/demineralised water connection 3/4"
Inlet water temperature 5–45 °C
Drain connection DN19
Water quality Pre-rinse with tap water of at least drinking water quality (observe drinking water regulations and local requirements). All further programme steps with demineralised or reverse osmosis water. This protects instruments and device from limescale and deposits.
Min. water inlet pressure 50 kPa (0.5 bar)
Max. water inlet pressure 1000 kPa (10 bar)
Spray arm rotation 25 to ≥ 50 rpm
Max. waste water temperature 93 °C
Waste water per hour approx. 20 l (in smaller intervals)
Drain pump capacity max. 15 l/min (volume in drain hose)
Length of inlet and drain hoses inlet hose 2.00 m; drain hose 2.50 m
Length of power cable 1.90 m
Internal memory up to 3000 cycles
Interfaces USB, network (RJ45)
Pollution degree (IEC 61010-1) category 2
Protection class (IEC 60529) IP20
CE marking 0297
Standards DIN EN ISO 15883-1/2, EN 61010-2-040
Flexio sieve basket 10 × 10 cm for small parts
Flexio sieve basket 10x10

The 10x10 mesh basket holds small parts, forceps and speculum holders so they can be reprocessed in line with requirements.

» About the product

Flexio instrument holder insert
Flexio instrument holder insert

Up to 72 standing instruments can be hygienically processed in the instrument holder insert. In addition, 8 additional and larger compartments offer space for vacuum cleaners and pliers.

» About the product

Bur stand with 4 cassettes including holder
Drill stand with 4 cassettes including holder

The Flexio drill stand 39 with a total of 156 slots is made of robust stainless steel. Our set contains four drill containers, each with 39 slots, which can be equipped directly in the individual treatment rooms, in...

» About the product

Flexio system carrier with 10 slots
Flexio system carrier 10 slots

The Flexio system carrier offers the option of attaching ten plug-in systems for rotating instruments or ear speculas.

» About the product

Flexio system carrier with 5 slots
Flexio system carrier 5 slots

The Flexio system carrier offers the option of attaching five plug-in systems for rotating instruments or ear speculas.

» About the product

Flexio plug-in system for 7 rotary instruments
Flexio plug-in system for 7 rotating instruments

The plug-in system for rotating instruments and mirror attachments can be integrated in a single version into the system carrier with 5 slots or in two versions in the system carrier with 10 slots.

» About the product

Flexio plug-in system for 4 ear funnels, loaded
Flexio plug-in system for 4 ear speculas

The plug-in system for ear specula can be integrated in a single version into the system carrier with 5 slots or in two versions in the system carrier with 10 slots.

» About the product

Flexio universal impression tray holder
Flexio Universal impression/spoon holder

The Flexio Universal impression/spoon holder is suitable for the hygienic preparation of impression trays, especially in the dental field, as well as for other medical instruments.

» About the product

Flexio holder for nasal specula and ear funnels
Flexio nasal specula/ear funnel holder

Up to 60 nasal speculas can be prepared in the nasal specula holder. You can also use it to hygienically process other medical instruments (up to 60 utensils).

» About the product

Flexio universal holder for 4 specula or 11 forceps, loaded with scissors
Flexio universal holder for 4 speculums or 11 pliers

The Flexio universal holder offers the possibility of preparing up to 4 speculas, 11 pliers or other medical instruments.

» About the product

Flexio universal holder for 8 specula or 21 forceps, loaded with scissors
Flexio universal holder for 8 speculums or 21 pliers

The Flexio universal holder offers the possibility of preparing up to 8 speculas, 21 pliers or other medical instruments.

» About the product

Flexio holder for small parts container, mono
Flexio holder for small parts container MONO

The Flexio holder MONO for holding a small parts container. The holder can be locked in both the basic basket and the 10x10 sieve basket. The Flexio holder can be equipped with a ball or multibox.

» About the product

Flexio holder for small parts container, duo
Flexio holder for small parts container DUO

The Flexio holder DUO for holding two small parts containers. The holder can be installed in both the basic basket and the 10x10 sieve basket. The Flexio holder can be equipped with a ball or multibox.

» About the product

Flexio small parts container, ball type, open and closed
Flexio small parts container ball

The fine-meshed Flexio small parts container "Ball" offers the option of storing small parts such as drills, matrices,...

» About the product

Flexio small parts containers
Multibox small parts container

The coarse-meshed Flexiro Multibox small parts container offers the option of storing small parts such as drills, matrices,…

» About the product

Flexio tray holder for 7 trays
Flexio tray holder 7

The Flexio tray holder offers the option of holding 7 trays. The trays may be a maximum of 4.5 cm wide.

» About the product

Safety data sheets
EC safety data sheet Oxides Thermotabs

(pdf ,457.79 KB) You will need the safety data sheet for the Oxides Thermotabs for your documents; please keep them carefully.

EC safety data sheet service tabs

(pdf ,357.62 KB) You will need the safety data sheet for the service tabs for your records; please keep them carefully.

Installation and instruction
Installation requirements for thermal disinfectors of the HYG 3 series V3.1

(pdf ,1.28 MB) Please check the installation requirements for your thermodisinfector carefully and send this document and signed order back to IC Medical.

Components
Functions
Technical Details
Additional Service
Instructions

Arrange your personal consultation appointment now!