“Do I really have to have my thermal disinfector validated?” – practices ask us this question regularly. The short answer is: yes. The longer answer explains why, what happens during validation and how to keep the effort low. This guide refers to the legal situation in Germany.

The legal basis in brief

  • German Medical Devices Operator Ordinance (MPBetreibV), Section 8: medical devices intended to be used with low microbial counts or sterile must be reprocessed using suitable validated procedures. The revised MPBetreibV has applied since 20 February 2025; reprocessing is still governed by Section 8.
  • Presumption of conformity: proper reprocessing is presumed if the joint recommendation of KRINKO (Robert Koch Institute) and BfArM “Hygiene requirements for the reprocessing of medical devices” is followed.
  • Infection Protection Act (IfSG), Section 23: practice owners must ensure that the measures required by the current state of medical science to prevent infections are taken.
  • DIN EN ISO 15883: the standard series for washer-disinfectors; validation follows the guideline of DGKH, DGSV and AKI.

This means: as the operator, you are responsible. The manufacturer supplies a suitable device – but you have to prove that your process works.

What exactly is validated?

It is not the device that is validated but the process: your thermal disinfector, your water, your process chemicals, your instruments and your loading patterns. Validation consists of three steps:

StepWhat is checked
Installation qualification (IQ)Is the device correctly installed and connected (power, water, waste water), and are all documents available?
Operational qualification (OQ)Does the device operate within the specified parameters – temperatures, holding times, dosing?
Performance qualification (PQ)Are your real loads reliably cleaned and disinfected? Checked with test pieces, temperature measurement and water analysis.

At the end you receive a validation report. It is your proof for the authorities – keep it safe.

Revalidation: not a one-off

A validated process must be checked regularly. In practice, performance requalification is usually carried out once a year; the validator sets the specific interval based on the results. In addition, revalidation may be required if something changes:

  • after a repair that may affect the process
  • when changing process chemicals
  • when the water quality or water treatment changes
  • with new instruments or loading patterns
  • after the device has been moved

What you have to do yourself in everyday operation

Validation does not replace routine. Between validations, you document and release every batch. This includes:

  • visual inspection of the instruments after each cycle
  • checking that the cycle was completed without errors (process parameters, batch record)
  • documented release by a competent person
  • regular maintenance according to the manufacturer’s specifications

The thermal disinfectors of the HYG series record batches automatically; documentation is done with the included IC Medical Printer Software.

Typical mistakes – and how to avoid them

  • Device bought, validation forgotten: many quotes do not include an initial validation. Ask about it.
  • Revalidation missed: without a reminder, the appointment is easily forgotten.
  • Chemicals changed without revalidation: changing the process chemicals changes the process.
  • Water quality underestimated: limescale and salts affect cleaning and instruments. Water treatment creates constant conditions.

How IC Medical supports you

  • Initial validation in the complete offer: for our thermal disinfectors HYG 3 and HYG 5, the initial validation by a qualified validation partner is included.
  • Independent validators: we work with accredited validation partners, so the manufacturer does not check itself.
  • Revalidation reminders: we remind you in good time and arrange the appointment.
  • Little variation in the process: a single programme, pre-dosed tabs instead of liquid chemicals and demineralised water ensure consistent conditions.
  • Advice from trained validators: our team includes trained validators (DGSV courses) – we advise you competently on validation and revalidation.

Frequently asked questions about validation

Does a thermal disinfector in a medical practice have to be validated?

Yes. Under Section 8 of the German Medical Devices Operator Ordinance (MPBetreibV), medical devices must be reprocessed using suitable validated procedures. What is validated is not the device alone but the reprocessing process in your practice – with your water, your loads and your process chemicals.

Is the manufacturer’s factory test sufficient?

No. Factory and type tests according to DIN EN ISO 15883 show that the device is suitable in principle. Validation proves that the process works reliably at the installation site with your instruments and loading patterns.

Who may validate a thermal disinfector?

Validation is carried out by qualified, independent validators with calibrated measuring equipment. IC Medical arranges experienced, accredited validation partners.

What happens if a thermal disinfector is not validated?

During an inspection by the competent authority, missing validation will be objected to. Depending on the severity, this can lead to requirements with a deadline, fines or, in extreme cases, a ban on reprocessing. In addition, you lack proof of proper reprocessing in the event of a liability claim.

May I reprocess manually instead of by machine?

According to the KRINKO/BfArM recommendation, machine cleaning and disinfection is required in principle for certain instruments and preferred for others. Manual procedures must also be standardised and validated – which takes considerable effort. A thermal disinfector makes validation much easier and reproducible.

When is revalidation required?

At regular intervals – in practice usually once a year – and after changes that may affect the process, for example after repairs, a change of process chemicals or a change in water quality.

Conclusion

Validation is mandatory – and not rocket science if it is planned from the start. When buying, make sure the initial validation is included, and organise revalidation reliably. Questions? Contact us or call +49 7181 70600.

Note: This guide provides a general overview of the situation in Germany (as of October 2026) and does not constitute legal advice. The applicable regulations and the requirements of your competent authority are decisive.